ELGA Labwater from UK is an integral part of Veolia Water Technologies, the world leader in water treatment.

ELGA specializes in delivering water purification systems for research, diagnostic, clinical and healthcare environments. ELGA has 75 years experience in water purification technologies. ELGA’s water purification system has unique features like continuous real-time water quality monitoring, recirculation of purified water stored in tanks, customizable audible and visual alarms etc.

3 types of water for every application in your lab

Ultrapure Water for analytical applications such as HPLC, cell culture, tissue culture or molecular biology techniques.

Pure Water for applications including general reagent preparation, buffers, pH solutions, histology and general chemistry.

 RO – permeate water for glassware rinsing, water baths, autoclaves, hydroponics and as feed water to Type 1 systems.

Key Features:

Ultrapure (Type I) water specifications (from fixed dispense tip)

Specifications Purelab Quest UV Purelab Quest
Resistivity 18.2 MΩ.cm @ 25°C 18.2 MΩ.cm @ 25°C
Dispense flow rate Up to 1.2 l/min (0.27 gal) Up to 1.2 l/min (0.27 gal)
TOC < 5 ppb < 30 ppb
Bacterial TVC < 0.1 cfu/ml *1 < 0.1 cfu/ml *1
Endotoxin < 0.001 Eu/ml *2 < 0.001 Eu/ml *2
RNases < 1 pg/ml n/a
DNases < 5 pg/ml n/a
pH Effectively Neutral Effectively Neutral
Particulates 0.2 μm filtration *1 0.2 μm filtration *1
Recommended daily volume Up to 10 l/day *3 Up to 10 l/day *3
Pure (Type II) water specifications (Water Outlet port 4)
Resistivity > 1 MΩ.cm @ 25 °C > 1 MΩ.cm @ 25 °C
TOC < 50 ppb < 50 ppb
Bacterial TVC < 100 cfu/ml < 100 cfu/ml
Recommended daily volume Up to 10 l/day *1 (2.2 gal) Up to 10 l/day *1 (2.2 gal)
RO-permeate (Type III) water specifications (Water Outlet port 5)
Conductivity < 20 μS/cm *⁴ < 20 μS/cm *⁴
TOC < 200 ppb *⁴ < 200 ppb *⁴
Bacterial TVC < 1000 cfu/ml *⁴ < 1000 cfu/ml *⁴
Ionic rejection > 96 % *⁵ > 96 % *⁵
Particulates and Bacteria rejection > 99% > 99%
Organic rejection (MW > 200 Da) 10l/hr *⁵ 10l/hr *⁵
Recommended daily volume Up to 30l/day Up to 30l/day

*1 When using point of use filters (LC134/LC197)

*2 When using point of use filter (LC197)

*3 Available volume of Type I and II water combined; increased use will reduce purification pack life

*4 Subject to suitable feed water purity (see ionic rejection) and system maintenance

*5 With feed water pressure at > 4 bar and temperature at 15 °C